Microprocessor-controlled Custom designed LCD with numerical presentation of twitch data
Automatic Calibration: Automatic adjustment for optimal gain and 100% reference Stimulation Modes: TOF (Train of Four) PTC (Post Tetanic Count) 1 Hz twitch 0.1 Hz twitch DBS 3.3 and 3.2 (Double burst) Tetanic stimulation (Burst), 5 sec - 50 Hz or 100Hz Output: (accuracy ± 5% of display value) Surface electrodes Constant current, 0 - 60 mA (0 - 12 µC up to 5 kOhm Monophasic, 200 µs pulse width Needle electrodes Constant current, 0 - 6 mA or (0 - 0.24 µC) Monophasic, 40 µs pulse width
Transducer: Acceleration transducer (accuracy ±5% of full scale value) Battery: 9V Alkaline or NiCd (6LR61 / 6AM6) Supply voltage: 9V=, 100 mA max Type: BF Weight: Approximately .55 lbs. with battery Dimensions: 8 x 3.6 x 2 inches
|